Bitte benutzen Sie diese Referenz, um auf diese Ressource zu verweisen:
doi:10.22028/D291-42711 | Titel: | Evaluating keratometry and corneal astigmatism data from biometers and anterior segment tomographers and mapping to reconstructed corneal astigmatism |
| VerfasserIn: | Langenbucher, Achim Taroni, Leonardo Coutinho, Catarina P Cayless, Alan Szentmáry, Nóra Hoffmann, Peter Wendelstein, Jascha Savini, Giacomo |
| Sprache: | Englisch |
| Titel: | Clinical & Experimental Ophthalmology |
| Bandnummer: | 52 |
| Heft: | 6 |
| Seiten: | 627-638 |
| Verlag/Plattform: | Wiley |
| Erscheinungsjahr: | 2024 |
| Freie Schlagwörter: | corneal astigmatism keratometry posterior cornea measurement toric intraocular lenses total corneal power |
| DDC-Sachgruppe: | 610 Medizin, Gesundheit |
| Dokumenttyp: | Journalartikel / Zeitschriftenartikel |
| Abstract: | Background: To compare results from different corneal astigmatism measurement instruments; to reconstruct corneal astigmatism from the postimplantation spectacle refraction and toric intraocular lens (IOL) power; and to derive models for mapping measured corneal astigmatism to reconstructed corneal astigmatism. Methods: Retrospective single centre study involving 150 eyes treated with a toric IOL (Alcon SN6AT, DFT or TFNT). Measurements included IOLMaster 700 keratometry (IOLMK) and total keratometry (IOLMTK), Pentacam kerato metry (PK) and total corneal refractive power in 3 and 4 mm zones (PTCRP3 and PTCRP4), and Aladdin keratometry (AK). Regression-based models map ping the measured C0 and C45 components (Alpin's method) to reconstructed corneal astigmatism were derived. Results: Mean C0 components were 0.50/0.59/0.51 dioptres (D) for IOLMK/ PK/AK; 0.2/0.26/0.31 D for IOLMTK/PTCRP3/PTCRP4; and 0.26 D for recon structed corneal astigmatism. All corresponding C45 components ranged around 0. The prediction models had main diagonal elements lower than 1 with some crosstalk between C0 and C45 (nonzero off-diagonal elements). Root-mean-squared residuals were 0.44/0.45/0.48/0.51/0.50/0.47 D for IOLMK/ IOLMTK/PK/PTCRP3/PTCRP4/AK. Conclusions: Results from the different modalities are not consistent. On average IOLMTK/PTCRP3/PTCRP4 match reconstructed corneal astigmatism, whereas IOLMK/PK/AK show systematic C0 offsets of around 0.25 D. IOLMTK/PTCRP3/PTCRP4. Prediction models can reduce but not fully eliminate residual astigmatism after toric IOL implantation. |
| DOI der Erstveröffentlichung: | 10.1111/ceo.14387 |
| URL der Erstveröffentlichung: | https://onlinelibrary.wiley.com/doi/10.1111/ceo.14387 |
| Link zu diesem Datensatz: | urn:nbn:de:bsz:291--ds-427119 hdl:20.500.11880/38305 http://dx.doi.org/10.22028/D291-42711 |
| ISSN: | 1442-9071 1442-6404 |
| Datum des Eintrags: | 28-Aug-2024 |
| Fakultät: | M - Medizinische Fakultät |
| Fachrichtung: | M - Augenheilkunde |
| Professur: | M - Univ.-Prof. Dr. Dipl.-Ing. Achim Langenbucher M - Prof. Dr. med. Nóra Szentmáry |
| Sammlung: | SciDok - Der Wissenschaftsserver der Universität des Saarlandes |
Dateien zu diesem Datensatz:
| Datei | Beschreibung | Größe | Format | |
|---|---|---|---|---|
| Clinical Exper Ophthalmology - 2024 - Langenbucher - Evaluating keratometry and corneal astigmatism data from biometers and.pdf | 2,82 MB | Adobe PDF | Öffnen/Anzeigen |
Diese Ressource wurde unter folgender Copyright-Bestimmung veröffentlicht: Lizenz von Creative Commons